Checking for Leaks 7
Agilent 7890 Series Troubleshooting 143
To Correct Leaks in the Split Splitless Inlet
If the inlet fails a pressure decay test, check the following:
Check the caps/plugs used in the test—make sure each is
correctly installed and tight.
If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
Check the tightness of the septum nut.
Check the septum. Replace if old or damaged.
Check the insert assembly installation.
Check the liner and liner O-ring. See To change the liner and
O-ring on the split/splitless inlet.
If you changed the gold seal, verify correct installation. See
To replace the gold seal on the split/splitless inlet.
Make sure the inlet temperature remained constant during
the test.
If these items do not resolve the problem, contact Agilent for
service.