30 Agilent 7890 Series Troubleshooting
2 ALS and Detector Symptoms
FID Does Not Ignite
Verify that the Lit Offset is 2.0 pA.
Ensure that the FID temperature is high enough for ignition
(>150 °C). Agilent recommends >300 °C.
Check that the FID ignitor glows during ignition sequence.
(See To Verify FID Ignitor Function During Ignition
Sequence.)
Check that the air and hydrogen pressures meet Agilent’s
recommendations (hydrogen > 35 psi [210 kPa] and air > 55
psi [380 kPa]). See the Agilent GC, GC/MS, and ALS Site
Preparation Guide.
Try increasing the supply pressures to the FID flow module.
This makes the flame easier to light without changing the
setpoints.
Increase hydrogen flow and decrease or turn off makeup gas
flow until ignition occurs, then reduce them toward the
method values. Experiment for the best values.
Increasing hydrogen flow and decreasing makeup flow will
help the FID ignite more easily. If it will light under these
modified conditions, the cause could be a partially clogged
jet, a weak ignitor, or a leak at the column fitting.
Check for a plugged or partially plugged jet. (See To Check
for a Plugged FID Jet.)
Measure the FID flow rates. Actual flow rates should be
within +/-10% of the setpoint. (See FID starting conditions.)
The hydrogen:air ratio greatly impacts ignition. Nonoptimal
flow settings can prevent flame ignition. (See To Measure a
Detector Flow.)
There could be a large leak in the system if the flame still will
not light. Large leaks result in measured flow rates being
different from actual flow rates, causing nonideal ignition
conditions. Thoroughly leak check the whole system,
especially the column fitting at the FID. (See Checking for
Leaks).
Check the column flow rate. (See To Measure a Column
Flow.) Hydrogen flow should be greater than the sum of the
column flow and makeup flow.
If the analysis permits, substitute nitrogen for helium as
makeup.

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals