66 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
3
Confirm that complete inlet maintenance has been
performed. This includes trimming and re-installing the
column.
4 Verify that the injecting device is working properly. If the
method uses a sampling valve, make a direct syringe
injection of concentration similar to that introduced by the
valve. Rule out a problem with the valve.
5 Inspect the mixing liner in the makeup gas adapter. A small
piece of graphite or sample contamination will reduce
sensitivity. See To Replace the uECD Fused Silica Indented
Mixing Liner and Install the Makeup Gas Adapter.
Contamination (high baseline)
If the ECD has a baseline higher than expected for its age, check
the following:
Rule out all other possible causes. See “Sensitivity” above.
Also check carrier and detector gas supplies, supply gas
traps, and the column.
Inspect the mixing liner in the makeup gas adapter. A small
piece of graphite can contaminate results. See To Replace the
uECD Fused Silica Indented Mixing Liner and Install the
Makeup Gas Adapter.
If the detector baseline is > 500 Hz, regardless of detector
age, bakeout the detector. Let it run for 1–2 days.

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals