Troubleshooting Tasks 8
Agilent 7890 Series Troubleshooting 205
To Adjust the FPD Lit Offset
To adjust the FPD Lit offset:
1 Press [Config].
2 Scroll to Front detector or Back detector (wherever the detector
is installed) and press [Enter].
3 Scroll to Lit offset. With the Lit offset line highlighted, enter the
new parameter for the FPD (typical value is 2.0 pA), and
press [Enter].