200 Agilent 7890 Series Troubleshooting
8 Troubleshooting Tasks
To Measure NPD Leakage Current
1 Load the analytical method.
2 Set the NPD Adjust Offset to Off and the Bead Voltage to 0.00 V.
Leave the NPD at operating temperature.
Leave flows on or off.
3 Press [Front Det] or [Back Det], then scroll to Output.
4 Verify that the output (leakage current) is stable and
< 1.0 pA.
5 The output should slowly drop toward 0.0 pA, and should
stabilize in the tenths of a picoamp. Current > 2.0pA
indicates a problem.
If the current is > 2.0pA, do the following:
Inspect the jet for contamination.
Check the column or column adapter fitting for leaks.
Ensure that the metal sealing ring is not crushed or
misaligned. See To Maintain the NPD Collector, Ceramic
Insulators, and Jet