EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7890A

Agilent Technologies 7890A Troubleshooting

Agilent Technologies 7890A
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #199 background imageLoading...
Page #199 background image
Troubleshooting Tasks 8
Agilent 7890 Series Troubleshooting 199
To Isolate the Cause of FID Noise
FID noise is the result of mechanical, electrical and chemical
factors. FID noise can be a subjective parameter. Often FID
baseline noise is perceived based on history of a given detector
or comparison with another detector in the lab. For proper
diagnosis of noise it is important to evaluate the detector noise
under documented conditions against a known standard. Find
more detailed information about noise in Noisy Detector,
Including Wander, Drift, and Baseline Spikes.
Before troubleshooting the detector, perform a noise test using
your Agilent data system. If the detector fails the noise test,
then troubleshoot the cause as described below.
To isolate the cause of FID noise:
1 If the noise test fails, remove the column and re-evaluate
detector noise with the FID capped and re-ignited, using only
H2/air and makeup detector gases. If it passes, suspect
contaminated column/carrier gas.
2 If the noise failed with no column installed, repeat the noise
test with only H2 and air - set makeup flow to "Off". If it
passes, suspect contaminated makeup gas.
3 If the noise test still fails, see To Measure FID Leakage
Current. If the leakage test fails, replace or clean the
collector and PTFE insulators, the interconnect with spring,
and/or the entire FID electrometer assembly. If the leakage
test fails, replace or clean the collector and PTFE insulators,
the interconnect with spring, and/or the entire FID
electrometer assembly.
4 If the leakage current test is OK, suspect a contaminated jet
or contaminated H2 or Air detector gas supplies (gases,
tubing, traps), especially if the background of the detector
when lit is >20 pA.

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals