EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7890A

Agilent Technologies 7890A Troubleshooting

Agilent Technologies 7890A
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
96 Agilent 7890 Series Troubleshooting
4 GC Not Ready Symptoms
A Gas Does Not Reach Setpoint Pressure or Flow
If an inlet does not reach its pressure setpoint, it will shut down
in an amount of time determined by the type of inlet. Do the
following:
Check for sufficient gas supply delivery pressure. The
pressure at the supply should be at least 10 psi greater than
the desired setpoint.
Check for leaks. (See “Checking for Leaks”.) A large leak may
be present somewhere in the system. Use an electronic leak
detector to find leaks, then correct them. Do not forget to
check the column—a broken column is a very large leak.
To troubleshoot a multimode or split/splitless inlet, see “To
Check for Leaks in a Multimode Inlet” or “To Check for Leaks
in a Split/Splitless Inlet”.
If using gas saver, be sure that the gas saver flow rate is high
enough to maintain the highest column-head pressure used
during a run.
Check for an incorrectly installed column.
Check for a defective inlet or detector pressure sensor.
If you are using a split/splitless inlet, MMI inlet, PTV inlet, or
volatiles interface:
Check the split ratio. Increase the amount of split flow.

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals