Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 85
NPD Low Selectivity
(High hydrocarbon response relative to nitrogen or
phosphorus.)
Verify that the hydrogen flow is correct (< 3 mL/min).
Inspect the bead; it may be defective or expended.
Verify correct bead voltage.
Replace the collector and insulators.