162 Agilent 7890 Series Troubleshooting
7 Checking for Leaks
1
Gather the following (see Consumables and parts for the PTV
inlet):
No-hole ferrule
1/4-inch wrench
Heat-resistant gloves (if the inlet is hot)
Column nut
New septum
New Graphpak 3D ferrule and liner
ECD/TCD Detector plug (part no. 5060-9055)
2 Load the inlet maintenance method and wait for the GC to
become ready.
3 Remove the column, if installed.
4 Plug the column fitting with a column nut and a no-hole
ferrule.
5 If using the septum head, and the quality of the septum (or
Microseal) and GRAPHPACK-3D ferrule on the glass liner are
unknown, replace them now. See To change the septum on
the PTV inlet and To change the liner on the PTV inlet.
6 Set the inlet to Split Mode.
7 Configure the column as 0 length.
8 Set the inlet temperature to 100 °C.
The test can find leaks at the: The test cannot find leaks at the:
septum column fitting
septum nut gas supply bulkhead fittings to the
flow module
liner O-ring seal tubing and connections in a transfer
line connected to the inlet
gold seal/washer and reducing nut
inlet body
flow manifold split vent valve
flow manifold septum purge valve
split vent tubing and trap
septum purge tubing
seals within the tubing between the
inlet flow module and the inlet body

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals