1 Installing the GC
52 Installation and First Startup
Procedure
1 Locate the input fitting for inlet coolant on the rear of the GC. Prepare enough tubing to
reach from the supply to this outlet.
2 Connect the supply tubing to the air supply outlet with the fitting recommended by the
supplier.
3 Use a Swagelok fitting to connect the supply tubing to the cryogenic valve input fitting.