EasyManua.ls Logo

Agilent Technologies 8890 - Making Swagelok Connections

Agilent Technologies 8890
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A Making Swagelok Connections
64 Installation and First Startup
Making Swagelok Connections
Objective
To make a tubing connection that does not leak and that can be taken apart without damaging
the fitting.
Materials needed
1/8-inch (or 1/4-inch, if used) preconditioned copper tubing
1/8-inch (or 1/4-inch, if used) Swagelok nuts
Front and back ferrules
Two 7/16-inch (for 1/8-inch nuts) or 9/16-inch (for 1/4-inch nuts) wrenches
Procedure
1 Place a Swagelok nut, back ferrule, and front ferrule on the tubing as shown in Figure 16.
Figure 16. Swagelok nuts and ferrules
2 Push the tubing into the stainless steel fitting.
3 Make sure that the front ferrule is fully engaged. Slide the Swagelok nut over the ferrule and
thread it onto the plug.
4 Push the tube fully into the plug, then withdraw it approximately 1 to 2 mm. See Figure 17.
Front ferrule
Back ferrule
Nut
Tubing
Important!
The narrow end of the
back ferrule fits into
the rear of the front
ferrule.

Table of Contents

Other manuals for Agilent Technologies 8890