EasyManuals Logo

Agilent Technologies 8890 Installation And First Startup

Agilent Technologies 8890
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
1 Installing the GC
Installation and First Startup 57
Step 21 Install the checkout column
A capillary column was shipped with the GC to be used to confirm proper operation. Agilent
suggests that it be used only for that purpose.
Before use, the column must be conditioned to remove any contaminants.
1 Locate the installation instructions for the column, inlet, and detector you will use. Refer to
the Maintaining Your GC manual. See to the sections on your specific inlet and detector
types:
2 Install the column in the inlet.
3 Connect the free end of the column to the detector, as described in Maintaining Your GC.
4 Turn on the carrier gas.
5 If using a split/splitless or multimode inlet, perform an Inlet Leak Check. Select Diagnostics
> Diagnostic Test, select Inlet #, then select Leak & Restriction Test. Press Start Test
to
begin the check. If the check fails, tighten connections.

Table of Contents

Other manuals for Agilent Technologies 8890

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8890 and is the answer not in the manual?

Agilent Technologies 8890 Specifications

General IconGeneral
BrandAgilent Technologies
Model8890
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals