A Making Swagelok Connections
Installation and First Startup 67
Using a Swagelok Tee
To supply gas from a single source to more than one input, use a Swagelok Tee.
NOTE
Do not combine valve actuator air with flame ionization air. The valve action will cause major
upsets in the detector signal.
Materials needed:
1/8-inch preconditioned copper tubing
Tubing cutter
1/8-inch Swagelok nuts and front and back ferrules
1/8-inch Swagelok Tee
Two 7/16-inch wrenches
1/8-inch Swagelok cap (optional)
1 Cut the tubing where you want to install the Tee. Connect the tubing and Tee with a
Swagelok fitting. See Figure 21.
Figure 21. Swagelok tee
2 Measure the distance from the Tee to the instrument fittings. Attach copper tubing to the
open Tee ends with Swagelok fittings.
Cap
Tee fitting
Nut and
ferrules