EasyManua.ls Logo

Agilent Technologies Agilent 8163A

Agilent Technologies Agilent 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Return Loss Measurement Return Loss Measurement
178 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
5 Move to [RLref] and make sure that the displayed value is correct. Set [RLref] to
the value of the return loss of the reference reflection you are using. For
example, if you are using the Agilent 81000BR reference reflector, set [RLref] to
0.18 dB.
Figure 95 Measuring the Reflection Reference
6 Switch on the Laser.
7 Press [Menu]. Move to <Reflectance calibration> and press Enter. The
instrument measures the power reflected by the back reflector. The [RL] value
changes to the same value as entered for [RLref].
Tip: An alternative step Step 7 is to select the Return Loss module’s
[Detail] screen, then press [RefCal]
Termination Calibration
Use this procedure to calibrate the Return Loss module against a
terminated cable, that is when there are no reflections returning from its
end.

Table of Contents

Related product manuals