Performance Test Instructions Performance Tests
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 355
Display/Key Functional Test
1 Setup the test equipment as shown in Figure 201 :
Insert the fixed laser source module in slot 1 and the power sensor in slot 2. If
you use the Agilent 8164A/B Lightwave Measurement System, insert the back-
loadable Tunable Laser module, if available, in slot 0.
2 Switch on the mainframe and wait until it has booted.
Testing Hardkeys
1 Press Config. You should see the Configuration menu.
2 Press Aux. You should hear abeep” and the Configuration menu should
disappear.
3 Press Appl. You should hear a “beep”.
4 Press Help. You should hear a “beep”.
5 Press Preset, the green hardkey. An egg-timer icon should be displayed while
all parameters are set to their default values. The Overview Screen showing all
channels should appear.
6 Press Channel several times.
The highlighted field on the display should toggle between the channels you
have inserted a module into.
Testing the Softkeys and the Cursor Key
1 Press the right arrow of the cursor key several times.
Check the fields on the display being highlighted one after another.
2 Press the left arrow on the cursor key several times.
Check the fields on the display being highlighted one after another.
3 Press the up arrow on the cursor key several times.
Check the fields on the display being highlighted one after another.
4 Press the down arrow on the cursor key several times.
Check the fields on the display being highlighted one after another
5 Using the cursor key, move to the power parameter, [P], of the Power Sensor
module.
Testing the Softkeys and the ENTER Key
1 Press [Menu]. The menu should appear.
2 Press [Close]. The menu should disappear.
3 Press [Details]. The details screen should appear. You should see the [Cal],
[Ref], and [Range] parameters in addition to [P] and [λ]. If you use the

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 8163A and is the answer not in the manual?

Agilent Technologies Agilent 8163A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 8163A
CategoryMultimeter
LanguageEnglish

Related product manuals