Cleaning Instructions How to clean bare fiber adapters
438 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
How to clean bare fiber adapters
Bare fiber adapters are difficult to clean. Protect from dust unless they are
in use.
Preferred Procedure
Use the following procedure on most occasions.
1 Blow away any dust or dirt with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the adapter:
1 Clean the adapter by pushing and pulling a new, dry pipe-cleaner into the
opening. Rotate the pipe-cleaner slowly as you do this.
2 Clean the adapter by rubbing a new, dry cotton-swab over the surface using a
small circular movement.
3 Blow away any remaining lint with compressed air.
Never use any kind of solvent when cleaning a bare fiber adapter as
solvents can damage the foam inside some adapters.
They can deposit dissolved dirt in the groove, which can then dirty
the surface of an inserted fiber.
CAUTION
Be careful when using pipe-cleaners, as the core and the bristles of
the pipe-cleaner are hard and can damage the adapter.
CAUTION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 8163A and is the answer not in the manual?

Agilent Technologies Agilent 8163A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 8163A
CategoryMultimeter
LanguageEnglish

Related product manuals