EasyManua.ls Logo

Agilent Technologies Agilent 8163A

Agilent Technologies Agilent 8163A
466 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and Maintenance GPIB Interface
312 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
GPIB Interface
You can connect your GPIB interface into a star network, a linear network,
or a combination star and linear network. The limitations imposed on this
network are as follows:
The total cable length cannot exceed 20 meters
The maximum cable length per device is 2 meters
No more than 15 devices may be interconnected on one bus.
Cables and Adapters
See GPIB Cables and Adapters” on page 333 for details on cables and
adapters.

Table of Contents

Related product manuals