Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
80
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
{<current>| MINimum | MAXimum}
This command programs the pending triggered current level of the power supply. The
pending triggered current level is a stored value that is transferred to the output
terminals when a trigger occurs. A pending triggered level is not affected by
subsequent CURRent commands.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
[MINimum | MAXimum]
This query returns the presently programmed triggered current level. If no triggered
level is programmed, the CURRent level is returned. CURRent :TRIGgered?
MAXimum and CURRent:TRIGgered? MINimum return the maximum and
minimum programmable triggered current levels.
VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<voltage>| MINimum | MAXimum}
This command directly programs the immediate voltage level of the power supply.
The immediate level is the voltage limit value of the selected output with the
INSTrument command.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
[MINimum | MAXimum]
This query returns the presently programmed voltage limit level of the selected output.
VOLTage? MAXimum and VOLTage? MINimum return the maximum and
minimum programmable voltage levels of the selected output.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{<voltage>| MINimum | MAXimum}
This command programs the pending triggered voltage level of the power supply. The
pending triggered voltage level is a stored value that is transferred to the output
terminals when a trigger occurs. A pending triggered level is not affected by
subsequent VOLTage commands.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
[MINimum | MAXimum]
This query returns the presently programmed triggered voltage level. If no triggered
level is programmed, the VOLTage level is returned. VOLTage:TRIGgered?
MAXimum and VOLTage:TRIGgered? MINimum return the maximum and
minimum programmable triggered voltage levels.

Table of Contents

Other manuals for Agilent Technologies E3631A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3631A and is the answer not in the manual?

Agilent Technologies E3631A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3631A
CategoryPower Supply
LanguageEnglish

Related product manuals