Chapter 4 Remote Interface Reference
System-Related Commands
84
System-Related Commands
DISPlay[:WINDow][:STATe] {OFF | ON}
This command turns the front-panel display off or on. When the display is turned off,
outputs are not sent to the display and all annunciators are disabled except the ERROR
annunciator.
The display state is automatically turned on when you return to the local mode. Press
the key to return to the local state from the remote interface.
DISPlay[:WINDow][:STATe]?
This command queries the front-panel display setting. Returns “0” (OFF) or “1” (ON).
DISPlay[:WINDow]:TEXT[:DATA] <quoted string>
This command displays a message on the front panel. The power supply will display
up to 12 characters in a message; any additional characters are truncated. Commas,
periods, and semicolons share a display space with the preceding character, and are
not considered individual characters.
DISPlay[:WINDow]:TEXT[:DATA]?
This command queries the message sent to the front panel and returns a quoted string.
DISPlay[:WINDow]:TEXT:CLEar
This command clears the message displayed on the front panel.
SYSTem:BEEPer[:IMMediate]
This command issues a single beep immediately.
Local

Table of Contents

Other manuals for Agilent Technologies E3631A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3631A and is the answer not in the manual?

Agilent Technologies E3631A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3631A
CategoryPower Supply
LanguageEnglish

Related product manuals