54 Chapter 4
Troubleshooting
Function Specific Troubleshooting
Figure 4-14 Bias Interface Simulator
Step 2. Set DC power supply output voltage +5 V. Connect TP2 (GND) on the bias interface
simulator to ‘-’ terminal of the power supply. Then connect TP1 (Vcc) on the simulator to
‘+’ terminal of the power supply. (refer to Figure
Figure 4-14 and Figure 4-15.)
NOTE DC power for the bias interface simulator can be supplied from the E4980A instead of
from an external DC power supply.
㪼㪋㪐㪏㪇㪸㫊㪼㪈㪇㪋㪍
(
(
1
(
(
1
㪪㪉
㪛㪠㪏㩷㪫㪦㩷㪈㪌
㪪㪈
㪛㪠㪇㩷㪫㪦㩷㪎
㪫㪧㪈
㪫㪧㪉