Chapter 4 55
Troubleshooting
Function Specific Troubleshooting
4. Troubleshooting
Step 3. Interconnect the bias interface simulator and bias interface connector on the rear panel of
the E4980A with the bias interface cable as shown in
Figure 4-15.
NOTE Connect using the adapter between the bias interface connector on the rear panel of the
E4980A and the bias interface cable, when you execute this step.
Figure 4-15 Bias Current Interface Function Test Setup
Step 4. Turn the E4980A ON.
Step 5. Press [System].
Step 6. Press the SELF TEST softkey to display the SELF TEST page.
Step 7. Use the CURSOR arrow keys to move the cursor to the TEST MENU field.
Step 8. Use the softkeys to select the BIAS INTERFACE test.
Step 9. Press the EXECUTE - TEST START softkeys to execute the BIAS INTERFACE test.
NOTE Check the settings of S1 and S2 described in the Step 1, if the E4980A’s LCD displays
“Illegal Test Setup”.
Step 10. Confirm the /RESET LED on the bias interface simulator turns ON.
Step 11. Confirm CS0 and CS1 LEDs on the simulator turns ON as the E4980A’s output signal is
Softkey Function
INCR + Increments the test number of the Test Menu.
DECR - Decrements the test number of the Test Menu.
㪼㪋㪐㪏㪇㪸㫊㪼㪈㪇㪋㪎
PrecisionLCRMeter
20Hz-2MHz
DC
DC
UNKNOWN
Discharge test device beforeconnecting
r42VPeakMaxOutput CAT I
DCSource
(Option 001)
Tri g ge r
DC Bias
DC
Sou rce
Ret urn
789
456
123
0
.
Pres et
Display
Format
Mea s
Set u p
Rec al l A Rec al l B
Save /
Rec al l
System
Local /
Lock
㪛㪚㩷㪧㪦㪮㪜㪩㩷㪪㪬㪧㪧㪣㪰
㪙㪠㪘㪪㩷㪠㪆㪝㩷㪪㪠㪤㪬㪣㪘㪫㪦㪩
㪜㪋㪐㪏㪇㪘
㪫㫆
㪙㪠㪘㪪㩷㪚㪬㪩㪩㪜㪥㪫
㪠㪥㪫㪜㪩㪝㪘㪚㪜
㪚㫆㫅㫅㪼㪺㫋㫆㫉
62)0&
628EE