GPRS Mobile Measurements
Measuring Output RF Spectrum
1. Connect the mobile.
2. Turn the mobile on and wait for Attached in the Active Cell: field.
Note: For mobiles that dont perform a GPRS attach automatically, set the mobile
to data mode.
3. Press the Start Data Connection (
F3) key and watch for the Active Cell:
field changing to Transferring.
4. Press the Measurement selection key.
5. Select the Output RF Spectrum measurement.
6. Press the ORFS Setup (
F1) key.
7. Press the Measurement Setup (
F1) key.
8. Set the measurement parameters as needed for your measurement, including:
Measurement Timeout = 20 s.
9. Press the Modulation Frequencies (
F2) key and set up your required offset
frequencies for ORFS due to modulation.
10.Press the Switching Frequencies (
F3) key and set up your required offset
frequencies for ORFS due to switching.
11.Press the Close Menu (F6) key.
12.Press the Return (
F6) key.
13.Press the Change View (
F2) key.
14.Press the Modulation Numeric 1 of 3 (
F1) key.

Table of Contents

Other manuals for Agilent Technologies E5515C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5515C and is the answer not in the manual?

Agilent Technologies E5515C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5515C
CategoryTest Equipment
LanguageEnglish

Related product manuals