GPRS Mobile Measurements
Measuring Bit Error
NOTE To make bit error measurements your mobile must be fitted with a Test
SIM card and have ETSI Type B connection capability.
1. Press the
CALL SETUP key.
2. Press the
More key (bottom right) to obtain the Call Parms (3 of 4) screen.
3. Press the Multislot Config (
F7) key and select 1 down, 1 up.
4. Press the Data Conn Type (
F4) key and select ETSI Type B.
5. Connect the mobile.
6. Turn the mobile on and wait for Attached in the Active Cell: field.
NOTE For mobiles that dont perform GPRS attach automatically, set the
mobile to data mode.
7. Press the Start Data Connection (
F3) key and watch for the Active Cell:
field changing to Transferring.
8. Press the
Measurement selection key.
9. Select the Bit Error measurement.
10.Press the Bit Error Setup (
F1) key.
11.Set the measurement parameters as needed for your measurement situation,
including:

Table of Contents

Other manuals for Agilent Technologies E5515C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5515C and is the answer not in the manual?

Agilent Technologies E5515C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5515C
CategoryTest Equipment
LanguageEnglish

Related product manuals