Using the Adv Fctn Key
Advanced Function Descriptions
53
PRESSURE VALVES
G1883 Adv Fctn, 9, Enter
Press Adv Fctn, 9, Enter to display the sampler’s pressure sensor readings for
either the vial pressurization gas or the carrier gas. Use the cursor keys to
scroll between the two choices. The display is read-only.
VIAL PRESS PSI XX (example)
CARR.PRESS PSI YY (example)
STABILIZE TIME
G1883 Adv Fctn, 6, Enter
G1289/G1290 Adv Fctn, 1, Enter
This function changes the length of time the oven equilibrates after reaching its
initial setpoint before vials are loaded into the oven. Selecting Stabilize Time
displays the following:
STABILIZE TIME 1.0
One minute is the default time. The allowed range of values is 0.0 to
99.9 minutes.
VALVE COUNT
G1883 Adv Fctn, ., 2, Enter
G1289/G1290 Adv Fctn, ., 2, Enter
Displays the number of times the sampling valve has actuated. Agilent service
personnel may refer to the valve count to get an idea of the extent of the
sampler’s use.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1289 and is the answer not in the manual?

Agilent Technologies G1289 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1289
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals