128
Error Messages
This chapter describes various error messages that accompany machine
failures and the means to correct the failures.
Most errors cause the sampler to stop running the current method. You may
need to remove vials from the carousel, re-load vials into the carousel, or
change the method if you want to continue your analysis from the point at
which it stopped.
Error Message Listing
Agilent 7694 or 7694 RESET
Agilent 7694E or 7694E RESET
Probable Causes
The sampler rebooted due to a power failure (or user input).
Comments
Power to the 7694 Headspace Sampler may be lost during a run. The sampler
contains a battery backup which will retain methods in storage. When power is
restored, the headspace sampler runs through a series of diagnostic tests. (See
“Power on Diagnostics on page 14 for more information.)
The method in use when the power was lost will be reloaded into the active
memory. If you press Start the sampler will run the method starting with the
first vial specified. If you wish to start the run from the point at which the run
was interrupted, you will have to edit the number entered for First Vial.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1289 and is the answer not in the manual?

Agilent Technologies G1289 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1289
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals