EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G1289

Agilent Technologies G1289 User Manual

Agilent Technologies G1289
157 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
Error Messages
Error Message Listing
129
ALARM STEPPER errors
Probable Causes
Carousel jammed because of mishandled vial or broken glass or other
foreign bodies.
Carousel out of alignment. See “Checking carousel adjustment” on
page 101.
Carousel drive mechanism failed.
Suggested Actions
1. Remove all vials from the carousel and gripper. Clean out the oven/
carousel. Power cycle the headspace sampler and hold down the clear key
until the tray is reset.
2. Call Agilent service.
ALARM TEMP1 (OVEN)
ALARM TEMP2 (LOOP)
ALARM TEMP3 (TUBE)
Probable Causes
The temperature sensor for the heated zone noted is failed, shorted, or
disconnected.
The temperature for the zone is too high, indicating a bad PC board or
sensor.
Oven > 220°C
Loop > 240°C
Tube > 240°C
Suggested Actions
Call Agilent service.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1289 and is the answer not in the manual?

Agilent Technologies G1289 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1289
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals