Troubleshooting
Mechanical/Electrical Problems
124
Check the stepper motor. Measure R across the left or right pair of wires in
the motor connector: R = 6–7 ohms.
Suggested Actions
1. Set agitation on with Vial Parameters.
2. Tighten or replace the belt in the oven assembly. See your service manual
or contact Agilent service.
3. Replace the stepper motor if necessary.
4. Call Agilent service.
Sample Tray Doesn’t Stop at Correct Location (G1289/G1290 only)
Probable Causes
Optical sensor is blocked or broken.
Faulty chip on the mother board.
Faulty encoder on the tray motor.
Suggested Actions
1. Make sure that the tab is in contact with the optical sensor.
2. Clean the sensor with an alcohol swab.
3. Call Agilent service.
Sampling Sequence Won’t Start
Probable Causes
The 7694 Headspace Sampler not ready.
Oven stabilization time not expired.
No vial in tray (G1289/G1290 only).
Setpoints not reached.
Suggested Actions
1. Check the 7694 Headspace Sampler setpoints.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1289 and is the answer not in the manual?

Agilent Technologies G1289 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1289
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals