Entering a Method
Storing and Loading Methods
64
Storing and Loading Methods
Storing a Method
Store Method places the method in active memory into one of four storage
locations for future use. To store the method currently in use:
1. Press Store Method. This displays the following message:
STORE METH. [1–4]
2. Enter the storage location desired {1–4} on the numeric keypad followed
by Enter.
The method currently in use is put into the specified memory location.
Parameters stored as parts of a method are Zone Temps, Event Times, Vial
Parameters, and the advanced functions for Oven Stabilization Time, Pressure
Units, and Multiple Headspace Extraction (G1289/G1290 only).
Caution Storing overwrites any method already in that location. There is no warning
message that an overwrite is taking place.
Note When the 7694 HS is first turned on after a loss of the battery–backed–up RAM,
the default method is loaded into all four method–storage locations.
Loading a Method
You can retrieve a method from memory for use in the current analysis by
using the Load Method key. To bring a method up from memory:
1. Press Load Method. The sampler displays the following message:
For the G1883, you can load method 5, the default method. This method
cannot be edited or chained.
2. Enter the number of the method you desire to retrieve on the numeric
keypad and press Enter.
G1289 and G1290 G1883
LOAD METH. [1–4] LOAD METH. [1–5]