The Keypad
Keys and Their Functions
35
where ### is the number of minutes from 1 to 999. If 0 is entered in the first
Shake screen, it will turn shaking off.
When a multi-vial method is started, the carousel shakes for the number of
minutes selected; then the first vial is lifted into the heating zone. After
injection, the vial immediately returns to the carousel.
If the vial heating time is shorter than the run cycle time, then shaking resumes
until the run cycle expires. Under these circumstances, the shaking time
setpoint only applies to the period before the first vial begins heating. The
additional shaking cannot be controlled or changed by the operator.
G1289/G1290
Sets the agitation level of the oven carousel. Agitation mixes the sample in the
vial for potentially faster equilibration and is used most often with liquid
samples. Settings are as follows: 0 = off, 1 = low, 2 = high. The stepper motor in
the oven rotates the carousel at a constant frequency to agitate the vials.
Agitation takes place during the entire vial equilibration time.
Load Method
Retrieves one of four stored methods and places it into active memory for use
in the current analysis. See “Loading a Method” on page 64.
Store Method
Stores the current method in one of four method storage areas. See “Storing a
Method” on page 64.
Caution Storing a method overwrites any method already stored there. There is no
warning message that an overwrite is taking place.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1289 and is the answer not in the manual?

Agilent Technologies G1289 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1289
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals