Entering a Method
Using a Method Chain for MHE
66
Deactivate a Method Chain
To deactivate a method chain:
1. Press Chain Methods. The display shows:
METH #+#+#+#
where # is the number of a method.
2. Press 0 (NO) to display:
METH CHAIN OFF
Note If Parameter Increment is active, Chain Methods is inactive. Turn off
Parameter Increment to use the method chain feature.
Using a Method Chain for MHE
Use Chain Methods to perform a multiple headspace extraction with multiple
punctures and constant equilibration time for each extraction.
To run an MHE:
1. Key in the parameters for Zone Temps, Event Times and Vial Parameters
for the method.
2. Store this method in one of the four storage locations.
3. Press the following:
Chain Methods {X + X + X + X }
where X is the location where your method is stored.
Using a Method Chain for Running More Than 44
Samples
By chaining the same method to itself, you can run up to 88 vials. To run more
than the 44 sample vials in the tray, do the following:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1289 and is the answer not in the manual?

Agilent Technologies G1289 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1289
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals