Using the Adv Fctn Key
Advanced Function Descriptions
44
Advanced Function Descriptions
CONSTANT HEATING TIME
G1883 Adv Fctn, 2, Enter
This function indicates that the 7694 is always in the constant heating time
mode.
DEFAULT METHOD
G1289/G1290 Adv Fctn, 7, Enter
Pressing Adv Fctn, 7, and Enter displays the following message:
LOAD DEFAULT METHOD
Pressing Enter again loads a default method into the headspace sampler. The
display shows the following flashing message:
METHOD LOADED
before returning to a top–level display. The following parameters are loaded in
the default method:
Oven Temp 70°C
Loop Temp 80°C
Tr. Line Temp 90°C
GC Cycle Time 25 min
Vial Eq.Time 15 min
Pressuriz.Time 0.2 min
Loop Fill Time 0.2 min
Loop Eq.Time 0.05 min
Inject Time 1.0 min
First Vial 1
Last Vial 10
Shake Off