Maintaining the Instrument
Installing the Transfer Line onto the Volatiles Interface
96
Installing the Transfer Line onto the Volatiles Interface
If you are using your headspace sampler with a 6890 and a volatiles interface,
install the transfer line onto the interface as follows:
1. Install an appropriate-sized male nut and steel ferrule onto the transfer
line.
2. Connect the transfer line to the interface. Tighten the nut 1/4-turn past
finger tight. Do not overtighten. If the nut leaks, tighten an additional
1/8-turn.
3. Check for leaks. See your 7694 Headspace Sampler Site Prep and
Installation Guide for details.
Transfer line
Nut
Volatiles
interface