Page #45 background imageLoading...
Page #45 background image
45
Installing the Autosampler
3
Installing the Thermostatted Autosampler
1 Place the Thermostat on the bench.
2 Remove the front cover and route the condensation drain tube to the waste.
Figure 15 Condensation Leak outlet
3 Remove the adhesive tape which covers the front door.
4 Remove the front door and remove the transport protection foam.
5 Place the Autosampler on top of the Thermostat. Make sure that the
Autosampler is correctly engaged in the Thermostat locks.
WARNING
Personal injury
To avoid personal injury, keep fingers away from the needle area during
Autosampler operation.
Do not attempt to insert or remove a vial or a plate when the needle is positioned.
WARNING
Damage through condensation
If the condensation tube is located in liquid the condensed water cannot flow out of
the tube and the outlet is blocked. Any further condensation will then remain in the
instrument. This may damage the instruments electronics.
Make sure that the condensation tube is always above the liquid level in the vessel.
8dcYZchVi^dcYgV^cijWZ
LVhiZWdiiaZ

Table of Contents

Other manuals for Agilent Technologies G1329A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1329A and is the answer not in the manual?

Agilent Technologies G1329A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1329A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals