14 G1369C LAN Interface User Manual
1 Introduction - Around your LAN Interface Card
Introduction to the LAN Interface Card
LAN Compatibility On Early 1100 Modules
All 1100 Series HPLC modules shipped prior to 1997 are NOT compatible
with the LAN Interface communication. The modules which host the LAN
Interface (usually the detector module) requires a new main board. The
serial number break of the 1100 series modules and the part numbers for
the new boards are listed below.
If an Agilent detector is part of the LC System, the LAN Interface Card should be inserted in
the detector module.
Table 3 LAN compatibility on early 1100 modules
1100 Module S/N break P/N Mainboard
G1310A below DE64300355, US64400233 G1311-66520 or higher
G1311A below DE64301137, US64401134 G1311-66520 or higher
G1312A below DE64300703, US64400425 G1312-66520 or higher
G1313A below DE64302092, US64400886 G1313-66520 or higher
G1314A below JP64201926 G1314-66521 or higher
G1315A below DE64301532, US64400333 G1315-66520 or higher