28 G1369C LAN Interface User Manual
2Getting Started
Automatic Configuration with Bootp
Automatic Configuration with Bootp
When automatic configuration with Bootp is selected and the LAN
Interface Card is powered on, it broadcasts a BOOTP (Bootstrap Protocol)
request that contains its MAC (hardware) address. A BOOTP server
daemon searches its database for a matching MAC address, and if
successful, sends the corresponding configuration parameters to the card
as a BOOTP reply. These parameters become the active TCP/IP parameters
immediately and the TCP/IP connection can be established.
Configuring the Agilent Bootp Service Program
NOTE
All examples shown in this chapter will not work in your environment.
You need your own IP- , Subnet- Mask- and Gateway addresses.
NOTE
Assure that the detector configuration switch is set properly. The setting
should be either Bootp or Bootp & Store, see Table 5 on page 22.
NOTE
Assure that the detector connected to the network is powered off.
NOTE
If the Agilent Bootp Service program is not already installed on your PC, then install it from
your Agilent ChemStation DVD, located in folder \Bootp.