Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 113
Error Information
6
Module Error Messages
Servo Restart Failed
Error ID: 2201, 2211
The pump motor in the module was unable to move into the correct position
for restarting.
When the module is switched on, the first step is to switch on the C phase of
the variable reluctance motor. The rotor should move to one of the C positions.
The C position is required for the servo to be able to take control of the phase
sequencing with the commutator. If the rotor is unable to move, or if the C
position cannot be reached, the error message is generated.
Probable cause Suggested actions
1
Mechanical blockage of the module.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2
Disconnected or defective cable.
Please contact your Agilent service
representative.
3
Blocked (passive or active) inlet valve.
Exchange the (passive or active) inlet valve.
4
Defective pump drive assembly.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7111A and is the answer not in the manual?

Agilent Technologies G7111A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7111A
CategoryWater Pump
LanguageEnglish

Related product manuals