Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 123
Test Functions and Calibration
7
System Pressure Test
System Pressure Test
Introduction
The System Pressure Test is used for checking the tightness of the LC system and
identifying leaks between the pump and a position in the flow path following
the pump blocked by a blank nut.
System requirements
Minimum software revisions:
Lab Advisor B.02.08
Minimum firmware revision: D.07.01.
Test Principle
A solvent can be chosen from available solvent channels and a maximum
pressure can be defined at which the test will be run. In contrast to older
revisions of this test, any solvent can be used.
Before the test, the pump and system are flushed with solvent in order to
remove air bubbles, as air bubbles are compressed during the test and
therefore would appear as leaks. Using a degasser is highly recommended.
Then the flow path is blocked by a blank nut at any position between the purge
valve and the column outlet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7111A and is the answer not in the manual?

Agilent Technologies G7111A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7111A
CategoryWater Pump
LanguageEnglish

Related product manuals