118 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
6
Error Information
Module Error Messages
Degasser: signal fail
Error ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser:
No valid pressure signal is shown during startup of the degasser
Measured pressure is higher than 32000 hPa
Degasser: Vacuum cannot be maintained
Error ID: 8014
Pressure in degasser vacuum chamber exceeds 180 mbar limit during normal
operation.
Probable cause Suggested actions
1
Degasser sensor defect
Please contact your Agilent service
representative.
2
Degasser sensor not connected to main
board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Liquid in degasser tubing.
Please contact your Agilent service
representative.
2
Leak in degasser tubing or chamber.
Please contact your Agilent service
representative.
3
Degasser vacuum pump defective.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7111A and is the answer not in the manual?

Agilent Technologies G7111A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7111A
CategoryWater Pump
LanguageEnglish

Related product manuals