Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 141
Maintenance
8
Checking and Replacing the Solvent Filter
Checking and Replacing the Solvent Filter
A functional solvent filter is essential for a good pump performance and for
protecting the LC system.
See “Bottle Head Assembly” on page 195 for related parts.
1 Remove the solvent filter from the inlet filter adapter and replace it by a
new one.
When If solvent filter is blocked.
Parts required p/n Description
5041-2168 Solvent inlet filter, 20 µm pore size
CAUTION
Small particles can permanently block the capillaries and valves of the module.
Damage of the module.
Always filter solvents.
Never use the module without solvent inlet filter.
NOTE
If the filter is in good condition the solvent will freely drip out of the solvent tube
(hydrostatic pressure). If the solvent filter is partly blocked only very little solvent will drip
out of the solvent tube.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7111A and is the answer not in the manual?

Agilent Technologies G7111A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7111A
CategoryWater Pump
LanguageEnglish

Related product manuals