Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 119
Error Information
6
Module Error Messages
Degasser: limit not reached
Error ID: 2244, 8053
Degasser is not ready within 8 min after startup, because the pressure inside
the vacuum chamber exceeds 120 mbar.
SSV Protection circuit triggered
Error ID: 2049
The protection circuit of SSV has switched OFF the valve. During
initialization, the processor checks the presence of the SSV connector. If a
defective valve or a wrong connection is detected the error message is
generated.
Probable cause Suggested actions
1
Liquid in degasser tubing.
Please contact your Agilent service
representative.
2
Leak in degasser tubing or chamber.
Please contact your Agilent service
representative.
3
Degasser vacuum pump defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Wrong valve connected.
Check if the correct valve is connected to the
correct port.
2
Defective Solvent Selection Valve.
Change the Solvent Selection Valve.
3
Defective main board
Contact your Agilent service representative.