EasyManuals Logo

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 109
Optimizing the Detector
5
Example: Optimization for Multiple Compounds
Select a Excitation
wavelength in the
low UV (230...260
nm). This will cover
nearly all
fluorescence in your
sample.
DO NOT select
additional emission
wavelengths (B, C,
D). Doing so will
increase the scan
time and will lower
the performance.
Figure 37 Detector Settings for Emission Scan

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals