Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 159
Error Information
7
Detector Error Messages
Flash Trigger Lost
Error ID: 6722
This message is displayed when the flash trigger is no longer generated.
Wavelength Calibration Failed
Error ID: 6703
This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the
message “Wavelength Calibration Failed” is displayed and the instrument stays in
a Not Ready condition.
Probable cause Suggested actions
1
Firmware problem.
Reboot the detector (power cycle).
2
Multi Mode Off
Please contact your Agilent service
representative.
3
Defective encoder.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Flash lamp not ignited or position not
correct.
Please contact your Agilent service
representative.
2
Cell position not correct.
Check the cell position.
3
Solvent in the cell not clean or air bubble in
the cell.
Flush the flow cell.
4
Monochromator assembly position not
correct (after replacement).
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals