Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 195
Test Functions and Calibration
8
Dark-Current Test
Test Failed
Probable cause Suggested actions
1
Defective PMT.
Exchange the PMT.
2
Defective reference diode or A/D converter.
Exchange the FLF board.