Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 15
Introduction to the Fluorescence Detector
1
How the Detector Operates
How the Detector Operates
Luminescence Detection
Luminescence, the emission of light, occurs when molecules change from an
excited state to their ground state. Molecules can be excited by different forms
of energy, each with its own excitation process. For example, when the
excitation energy is light, the process is called photoluminescence.
In basic cases, the emission of light is the reverse of absorption, see Figure 3
on page 15. With sodium vapor, for example, the absorption and emission
spectra are a single line at the same wavelength. The absorption and emission
spectra of organic molecules in solution produce bands instead of lines.
Figure 3 Absorption of Light Versus Emission of Light
DEVRUSWLRQ
HQHUJ\OHYHO
HQHUJ\OHYHO
OXPLQHVFHQFH
HQHUJ\OHYHO
HQHUJ\OHYHO
K
K
ν
ν

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals