EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>G7121B

Agilent Technologies G7121B User Manual

Agilent Technologies G7121B
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #209 background imageLoading...
Page #209 background image
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 209
Maintenance
9
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Table 28 Simple repairs
Procedure Typical Frequency Notes
Flow cell exchange If application requires a different flow cell type or if
defective.
Complete Assembly
A wavelength calibration check should
be performed after replacement.
If the flow cell is removed and inserted,
then a quick calibration check is
performed.
If this fails, you must do a wavelength
recalibration, see “Wavelength
Verification and Calibration” on
page 182.
Flow cell flushing If flow cell is contaminated.
Leak sensor drying If leak has occurred. Check for leaks.
Leak handling System
replacement
If broken or corroded. Check for leaks.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals