172 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Lamp Intensity Test
Lamp Intensity History
Results of the lamp intensity test (if the last one is older than one week) are
stored as lamp history (date code, intensity of four different wavelengths,
250 nm, 350 nm, 450 and 600 nm) in a buffer. The data/plot can be retrieved
via the diagnostics and provides intensity data over a length of time.
Figure 58 Lamp Intensity History (Agilent Lab Advisor under Module Info)