176 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Raman ASTM Signal-to-Noise Test
Using the Agilent Lab Advisor
1 Set up the HPLC system and the Lab Advisor.
2 Flush the flow cell with clean bi-distilled water.
3 Start the test in the Lab Advisor.
Figure 61 Raman ASTM Signal-to-Noise Test (Agilent Lab Advisor)
In case of failing this test (as shown above) see “Interpretation of the
Results” on page 177.