130 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
5
Optimizing the Detector
Changing the Xenon Flash Lamp Frequency
Changing the Xenon Flash Lamp Frequency
Modes
The lamp flash frequency can be changed into the following modes:
Best sensitivity can be expected with “no economy”, see Figure 48 on page 130.
Figure 48 Xenon Flash Lamp Frequency
Table 17 Flash Lamp Modes
Positioning 296 Hz (Standard), 560 V 63 mJ (18.8 W)
74 Hz (Economy), 560 V 63 mJ (4.7 W)
Rotation (Multi Ex/Em) 74 Hz (Standard), 950 V 180 mJ (13.3 W)
74 Hz (Economy), 560 V 63 mJ (4.7 W)
6WDQGDUG +]