160 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
7
Error Information
Detector Error Messages
Wavelength Calibration Lost
Error ID: 6691
After exchanging the monochromator assemblies, the calibration factors
should be reset to defaults values (a new FLM board comes with default
values). In this case “Wavelength Calibration Lost” is displayed and the
instrument stays in a Not Ready condition.
Flow Cell Removed
Error ID: 6616, 6702, 6760
The detector has an automatic cell recognition system. When the flow cell is
removed, the lamp is turned off and a NOT READY condition exists. If the flow
cell is removed during an analysis, a SHUT DOWN is generated.
Probable cause Suggested actions
1
Reset of monochromator settings after
exchange.
Perform a wavelength calibration.
2
Replacement of FLM board.
Perform a wavelength calibration.
Probable cause Suggested actions
1
Flow cell has been removed during analysis.
Insert flow cell and turn on the lamp.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals