Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 27
Introduction to the Fluorescence Detector
1
Analytical Information From Primary Data
Analytical Information From Primary Data
We now know how the primary data from your sample is acquired in the
optical unit. But how can the data be used as information in analytical
chemistry? Depending on the chemistry of your application, the luminescence
measured by the fluorescence detector will have different characteristics. You
must decide, using your knowledge of the sample, what mode of detection you
will use.
Fluorescence Detection
When the lamp flashes, the fluorescing compounds in your sample will
luminesce almost simultaneously, see Figure 13 on page 27. The luminescence
is short-lived, therefore the fluorescence detector need only measure over a
short period of time after the lamp has flashed.
Figure 13 Measurement of Fluorescence
7UDFNDQG+ROG
7LPH>VHF@
,QWHQVLW\
,JQLWH

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121B and is the answer not in the manual?

Agilent Technologies G7121B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals